SCFvh 4002 Perfection Laboratory fridge-freezer with NoFrost

Exterior dimensions: height/width/depth 204.4 / 59.7 / 65.4 cm
Adjustable temperature range +3 °C to +16 °C
Energy consumption per year 277 kWh/a

DIN 13277 compliant

DIN 13277 compliant

Laboratory and pharmacy refrigerators from Liebherr meet all the necessary features and performance criteria to fully comply with DIN 13277. DIN 13277 defines the requirements and test specifications for refrigerators and freezers used in laboratories and for storing medicinal products.

+2 °C SafetyDevice

+2 °C SafetyDevice

If the temperature falls below +2 °C due to a malfunction while the temperature setting is at +5 °C or higher, the SafetyDevice takes control, stabilising the temperature, activating the alarm and protecting the contents from freezing. At colder settings, this activation threshold decreases relatively with the temperature setting.

Ultraefficient

Ultraefficient

Energy-efficient refrigerators and freezers help with ever increasing electricity prices. Liebherr fully utilises the savings potential of every component – from the compressor to the seal, to the lighting and electronics. The ultra-efficient appliances in the Perfection series for example demonstrates how this has a positive effect on your overall operating costs.

NoFrost technology

NoFrost technology

With professional laboratory appliances from Liebherr, defrosting is no longer necessary. Any moisture that occurs is bound by the fan-assisted cooling, and needs-based defrosting phases drive even the last moisture particles out of the interior. For convenient, safe, and ice-free storage.

Mechanical lock

Mechanical lock

All professional Liebherr appliances in the Performance series are equipped with a mechanical lock with two keys for the storage of temperature-sensitive products. This way you can be sure that no unauthorised persons can access your stored goods.

Easy to clean interior

Easy to clean interior

The plastic inside container is made of polystyrene: a fully recyclable, robust, and odourless plastic. Manufactured seamlessly from one piece, it has no hidden dirt traps and the large corners can be cleaned quickly and easily. The perfect combination of sustainability and hygiene.

Alarm relay

Alarm relay

In the event of an alarm, the responsible control centre is informed immediately, usually through connection to the building control system. You can define in advance which alarms are to be forwarded, as well as set how long the forwarding remains active and whether a reminder should be sent after the alarm acknowledgement. This allows you to react immediately in critical situations.

Additional access port for calibration

Additional access port for calibration

An additional feedthrough can be retrofitted for regular calibrations or measurements. For this purpose, a hole is drilled for a cable inlet with a maximum diameter of 40 mm at a specially designated location.

Accurate 1-point calibration

Accurate 1-point calibration

The laboratory appliances with electronic controls have a 1-point calibration function for extremely accurate temperature control. This permits compensation between the set temperature and the actual internal temperature. The correction value can be adjusted in increments of 0.1 K.

Intuitive menu navigation

Intuitive menu navigation

Self-explanatory, convenient, and clear: Seamlessly integrated into the front, the control panel shows detailed appliance information and your current appliance status on a large display, which is colour-coded according to the category. The menu navigation is self-explanatory and is controlled by “touch and swipe”.

Antibacterial lever handle

Antibacterial lever handle

A benefit for users but a bane for germs: The easy-turn handle mechanism makes opening the door effortless. It contains special antimicrobial substances that reduce bacteria by up to 99.99%*. Regular and time-consuming disinfection is not necessary. This means the appliance is extremely safe even if many people have access to it. *According to ISO 22196:2011

Clear appliance information

Clear appliance information

All of the information about the appliance should be there exactly when you need it. That's why every Liebherr has its serial and service number as well as the model designation on the display, where it can be easily read. Frustrated searches or even having to clear the interior out to be able to read the information are now a thing of the past.

Interchangeable door hinge

Interchangeable door hinge

Liebherr appliances are fitted with a right-hand door hinge as standard. Without much effort – and without any additional parts – the door hinge can be changed to the left. This allows you maximum flexibility when choosing your installation location.

Digital temperature display

Digital temperature display

The digital temperature display indicates the refrigerator’s interior temperature to the exact degree. This can be easily read from the outside to give users immediate information and without the need to open the appliance.

Evaporator cover

Evaporator cover

The evaporator is secured with a cover to prevent stored substances from coming into contact with its cold surface. This means that it is impossible for substances to accidentally freeze on its surface. In addition, the evaporator cover ensures even greater temperature stability.

LED front lighting

LED front lighting

The LED front lighting provides optimal illumination of the drawers even in dark environments. Thanks to the perfect overview, the desired substance is immediately at hand, the door can be quickly closed again, and the interior temperature remains stable. For low consumption and lower energy costs.

Vertical LED lighting

Vertical LED lighting

Vertical LED lighting provides optimal illumination of the interior over the entire height of the appliance. Thanks to the perfect overview on all levels, the desired substance is immediately at hand, the door can be quickly closed again, and the interior temperature remains stable. This saves you unnecessarily searching for something – and above all energy costs.

Reminder function

Reminder function

The interval timer reminds you of the date of the next basic cleaning or gives the weekly reminder to check the expiration dates of the products. The interval can be selected from predefined intervals and set directly on the appliance. Leaving you to focus solely on your work.

Power failure alarm

Power failure alarm

If a malfunction occurs, an integrated battery provides a visual and audible alarm and takes over the power supply. As soon as the power is restored, the time of failure and the maximum temperature reached are displayed.

SafetyStop

SafetyStop

The grid shelf with the pull-out stop ensures stability and a safe working environment: The grid shelf cannot be pulled out by mistake – even if things start to get a little hectic. When stocking or emptying the appliance, it remains reliably fixed in position – and you are spared any mishaps.

Self-closing door < 90° / automatic door-closing at less than 90°

Self-closing door < 90° / automatic door-closing at less than 90°

Carry out routine work in comfort; reassurance nothing will be missed: From an opening angle of approx. 90°, the appliance door remains open and facilitates stocking and unloading as well as cleaning. Below approx. 90°, it closes by itself so that the appliance does not remain open accidentally. This keeps the temperature stable and preserves the quality of your groceries.

Settings with password protection

Settings with password protection

All settings that influence the function or control of the appliance can be secured by password. This means that the contents are optimally protected from unwanted changes or manipulation to the storage conditions.

Ten years spare parts availability

Ten years spare parts availability

Sustainable means the appliance is not only environmentally friendly and economical, but also durable. To ensure that you can enjoy your Liebherr appliance for a long time, Liebherr guarantees spare parts availability of at least ten years. Although we do everything we can to ensure that spare parts are not needed in the first place.

Height-adjustable feet

Height-adjustable feet

With height-adjustable feet as an accessory, the height of the appliance can be adjusted to your needs. For example, you could raise your professional appliance to clean under it more easily.

Alarm history

Alarm history

In order to be able to evaluate the quality of the stored substances, a clear history of anomalies that have occurred is necessary. The alarm history records important data about the last ten alarms such as date, duration or maximum temperature. This way, you don't rely on your memory to perform the check but rather rely on the facts.

Alarm test

Alarm test

An alarm is only ever any good if it can be triggered. The alarm test can be used to check the functionality of the internal and possibly an externally connected alarm device. The cooling of the appliance is not interrupted during the test run, so you can arrange an emergency simulation and rehearse with your team as you please.

Calibration function

Calibration function

Due to operational requirements and application-specific conditions, it is possible to deviate the temperature of the interior from the set temperature to a storage temperature determined with calibrated thermometers. These deviations can be adjusted by the calibration function. For the secure feeling of cooling at precise temperatures.

Replaceable door seal

Replaceable door seal

Save time and money: Liebherr's plug-in door seals can be easily replaced. If a seal is damaged, you can replace it yourself without any tools.

Cleanroom compatible

Cleanroom compatible

Liebherr fridges and freezers in the new scientific line have been tested for use in cleanrooms in accordance with DIN EN ISO 14644. As a result, an additional risk assessment by the customer to determine cleanroom suitability is no longer necessary. This saves time and money.

Demo mode

Demo mode

For training purposes or exhibitions, the functionalities of the appliance can be simulated in demo mode without having to operate the appliance. This saves time and money as the cooling function is not needed.

Feedthrough for external sensors

Feedthrough for external sensors

To additionally control the storage temperature with the help of a separate sensor, wiring from the outside to the inside is required. The cable feeding offers a clean and professional solution here. The integrated opening allows connection without drilling holes or breaking the door seal.

Hidden customer menu

Hidden customer menu

Protected, secure, simple: The hidden customer menu provides users with status information about the appliance or allows them to make detailed appliance settings, such as resetting to factory settings. Since this information and the settings are located on a separate, code-protected menu level, unwanted changes or manipulation are avoided.

Hidden technician menu

Hidden technician menu

The hidden service menu allows individual configuration of the appliance, such as alarm sound on/off, adjustment of target value ranges or changes to automatic defrosting. By making technical fine adjustments, the appliance can be adapted to your specific needs.

Integrated data logger

Integrated data logger

To ensure full traceability independent of the monitoring systems, the integrated data logger records the temperature profile, alarm history, and service information for up to five years. The stored data can be read out directly on the appliance via a USB interface.

LightAlarm

LightAlarm

In case of an uncontrolled temperature rise, if a door has been left open or there were any other problems, the interior light of the appliance will start flashing. In the case of glass-door appliances, this quickly reveals where action is needed. In the case of solid-door appliances, the flashing interior warns to close the door immediately in order to preserve the storage quality.

Manual defrost start

Manual defrost start

Doors left open or evaporating substances can cause icing on the evaporator resulting in temperature fluctuations and higher energy consumption. Via menu selection, you can start an automatic defrost to clear the appliance of ice, restoring optimal storage conditions and lower energy consumption.

Almost maintenance free

Almost maintenance free

Saves time, energy, and money: Liebherr’s professional freestanding appliances operate without filters or other components and parts that require regular replacement. Only cleaning and calibration in line with your individual needs are necessary on a regular basis.

IEC 61010-2-011 Standard

IEC 61010-2-011 Standard

Liebherr refrigerators and freezers specially developed for storing temperature-sensitive products are CE safety-compliant. This ensures that all mechanical and electrical requirements for injury-free operation are met – and that you and your team are kept safe.

Product sensor PT1000

Product sensor PT1000

In addition to the standard control sensor, a separate PT1000 sensor is connected to measure a simulated product temperature. The product sensor temperature shown in the display can be calibrated independently, instrument alarms and min./max. limits refer to this value. For greater storage reliability based on real product values.

Quick start guide

Quick start guide

No one likes to read the instructions. That's why Liebherr offers the most important information at a glance on two pages. This includes information on menu navigation, the electronic lock, cleaning, and general topics such as stored goods that are or are not permitted. Leave the technician in charge of reading the original manual.

Recommended cleaning products and disinfectant

Recommended cleaning products and disinfectant

Liebherr appliances are designed to have a product lifetime of 15 years. To maintain this, cleaning and disinfecting agents that have been tested by Liebherr for safe use are recommended separately, or in the manual. This way you can be sure that you are not harming your appliance by mistake when you are actually trying to take care of it.

Min./max. temperature recording

Min./max. temperature recording

Always keep an eye on the temperature: Min./max. temperature recording stores the minimum and maximum temperature values of the interior that have occurred from a defined start time and shows them clearly in the display. This way you can be sure that, up until the moment you read off the values, the storage temperatures were always within the tolerated limits.

Stainless steel workmanship

Stainless steel workmanship

The interior, shelves and inner doors are made of rust-proof stainless steel and are therefore easy to clean. This enables optimum hygiene.

Model type Laboratory fridge-freezer with NoFrost
Classification Perfection
GTIN 9005382247594
Energy consumption per year 277 kWh/a
Energy consumption in 24h 0.757 kWh / 24h
Noise level 48 dB
Ambient temperature +10 °C to 35 °C
Refrigerant R 600a
Heat emission 290 kJ / h
Heat distribution system Air cooling
Rated power in watts (Catalog) 140.0 W
Voltage 220-240 V ~
Frequency 50/60 Hz
Control unit Touch & Swipe
Power failure alarm immediately upon power failure for min. 12h
Malfunction: Warning signal optical and acoustical
positive temperature deviation from the setpoint 1.2 Kelvin
negative temperature deviation from the setpoint 2.0 Kelvin
Increase in temperature from +5 °C to +10 °C (empty, +25 °C TU) 70 min
Increase in temperature from -20 °C to -15 °C (empty, +25 °C TU) 50 min
Decrease in temperature from +25 °C to +5 °C (empty, +25 °C TU) 30 min
Cooling time from +25 °C to -20 °C (empty, +25 °C TU) 150 min
Recovery time after 1 min of door opening (empty, +25 °C TU, according to EN 60068-3) 10 min
potential-free contact
SmartMonitoring-enabled Yes
Connectivity type SmartModule
Connectivity solution integrated, removable
Interface WLAN/LAN
Data logger integrated, USB interface
min./max. temperature recording Yes
Alarm history Yes
Product sensor integrated
Gross volume cooling 272 l
Net volume cooling 206 l
Adjustable temperature range +3 °C to +16 °C
Temperature display external digital
Cooling system, refrigerator compartment dynamic
Defrosting method automatic
Interior light LED light column, left
Number of storage shelves 5
of which adjustable 4
Bottom grid shelf
Gross volume freezing 112 l
Net volume freezing 58 l
Adjustable temperature range -9 °C to -30 °C
Temperature display external digital
Cooling technology NoFrost
Cooling system, freezer compartment dynamic
Defrosting method automatic
Interior light LED light
Drawers in freezer compartment 3
Number of storage shelves 3
of which adjustable 0
Bottom grid shelf
Side wall material steel
Colour White
Door/Cover material Steel
Handle Antimicrobial handle with opening mechanism
Material of interior containers Plastic white
Storage shelf material, refrigerator compartment Plastic-coated grids
Material of adjustable shelves in the freezer compartment Glass
Material of adjustable feet Steel, zinced
Self-closing door
Door hinges Right
Door hinge replacement Possible for customer
Door seal Replaceable
Height-adjustable feet — mm
Castors
Type of castors
Type of lock mechanical
Max. installation altitude 2,000 m above sea level
Connector cable (length) 3,000 mm
Total gross volume 384 l
Total net volume 264 l
Exterior dimensions: height/width/depth 204.4 / 59.7 / 65.4 cm
Height / Width / Depth (with packaging) 2,080.0 / 615.0 / 725.0 mm
Weight (without packaging) 83 kg
Weight (with packaging) 90 kg
Load-bearing capacity of shelf areas, refrigerator compartment 45 kg
Load-bearing capacity of shelf areas, freezer compartment 40 kg
Net width of shelves 46 cm
net depth of storage shelves 42.3 cm
Certifications DIN 13277 / IEC 60068-3 / IEC 61010-1 / IEC 61010-2-011 / IEC 61326-1
Cleanroom class according to ISO EN 14664-1 ISO class 5

Please note: despite careful data management, we reserve the right to make technical changes; and discrepancies may occur between the image or text content and the original appliance. *According to guarantee terms and conditions for the 10-year manufacturer’s guarantee: home.liebherr.com/guarantee (valid only in UK and IRL)